HOLIDAY HOURS: Our facilities will be closed on December 24th and 25th in observance of the Christmas holiday. Normal business hours will resume on December 26th.

Skip To Content

Quantim® QMC Coriolis Mass Flow Meters & Controllers

New third-generation Quantim® QMC Coriolis flow meters are ideal for very low-flow gas and liquid measurement and control applications.
Models QMC / QMB

Achieve the most accurate, stable, repeatable and reproducible measurement and control of very low flow rates of liquid or gas with the new series of Quantim® QMC Coriolis flow meters and controllers from Brooks Instrument.

This third-generation Quantim® QMC Coriolis flow meter is engineered to improve process quality, production yield and throughput and reduce scale-up time from lab to pilot plant to production. The devices use a modern IoT-equipped digital architecture that, combined with the next-generation sensor design, enables faster response time with little to no overshoot when adjusting gas or fluid flow, improved zero stability and expanded I/O options.

In addition, Quantim® QMC Coriolis flow controllers provide multivariable outputs, including mass flow, volume flow, density and temperature. They also include a suite of advanced alarms and in situ diagnostics that actively monitor the process to identify system issues earlier and alert operators when mass flow controller maintenance or calibration is required.

Questions?

Talk to one of our application engineering experts by calling: 1-888-554-3569


For pricing information:

Request a Quote
Quantim® Series Data Sheet

Features

  • True mass flow measurement
  • Accurate density measurement
  • NEW I/Os: 0-10 Vdc, 0-20 mA, RS485 S-Protocol
  • Variety of material options, enclosure types and area classifications available
  • NEW easy-to-access external service port
  • NEW Quantim® QMC Coriolis series is form-, fit-, and function- compatible with prior Brooks Instrument Quantim® models

Benefits

  • Process chemistry and/or process conditions can be altered without the need to change or recalibrate the measurement system
  • With an all-stainless steel construction, Quantim® QMC is compatible with most process fluids
  • Accessible via a 2.5 mm jack plug, the external service port coupled with our free downloadable Brooks Expert Support Tool (BEST) software, facilitates enabling the on-board alarms and diagnostics for enhanced process monitoring and predictive maintenance, simplifies installation, supports troubleshooting and calibration verification

Applications

{5D094FEF-32CD-4838-93A9-DA0CA6E5123C}
Product Type
Mass Flow Controller
{5003CEFC-F205-4FDC-BA26-1DB1A7A9E454}
Differentiator
Third generation Coriolis sensor and electronics provides superior performance and reliability
{727CCEF2-64E1-4F75-ADDF-839A2874A3F9}
Performance
{3C1055DF-0814-4CF6-92F6-30795FE269CC}
Repeatability
±0.05% or ±[0.5 x (zero stability/flowrate) x 100]% of rate whichever is greater
{A742E6D4-0F87-4F01-8FE9-2F40266B5836}
Response Time

SS Controller: <2 sec

SS Meter: <0.5 sec

{C5082484-2CE7-415F-925A-42B948747789}
Flow Range (Full Scale Capacity)
10 - 27,000 gram/hr
{66BC80C1-5F4D-4961-82EC-E6A4286F27DE}
Temperature Accuracy
± 0.5°C
{AE9454C9-4B0B-45E2-AB08-283AEF2DB4C8}
Accuracy

Standard Flow Accuracy (below) or [(zero stability/flow rate) x 100]% of rate, whichever is greater

Liquid: 0.2% of rate

Gas: 0.5% of rate

{3580C4C8-0781-40BF-94AF-227353DB0BF6}
Differential Pressure Range

Liquid: 10 - 200 psi

Gas: 10 - 150 psi

{2EA51D7A-6757-418D-963F-2DB8DAB3D730}
Density Range
0 to 2.0 g/cc
{65B1F0E6-0156-43A5-B5D2-EBB9D70F5FAD}
Density Accuracy
± 0.005 g/cc
{B76BF611-2FB2-42D5-A875-E1A6D7C3F09C}
Leak Integrity

Elastomer Seal: Outboard 1 x 10-9 atm. cc/sec, helium (max)

Metal Seal: 1 x 10-10 atm. cc/sec, helium (max)

{96ABEBA1-1E72-4B50-9029-0466EAC48B9D}
Mechanical
{FE84826B-179F-4454-BEAF-225EA6274174}
Wetted Materials

316L, 316L VAR, High Alloy Ferritic Stainless & 17-7PH

{7355A946-A120-4787-8254-B43E198D2655}
Seal Material
Viton, Buna, Kalrez, EPDM or Metal
{2AB126DD-F30B-4D74-B5A4-7EF814CE2DEB}
Housing

Standard: IP40

{BB2325C7-C4C6-49DF-A187-57F255710BE7}
Inlet Filter

Size 2 Sensor: 1 & 10 micron

Size 3 & 4 Sensors: 10, 20, 30 & 40 micron

{BEF690DA-1E53-4FFD-8834-931BDA4A79E4}
Fittings

Elastomer Seal: 1/16”, 1/8”, 1/4” or 6mm tube compression

1/8” or 1/4” NPT(F)

1/8” or 1/4” VCR, 1/4” VCO

Metal Seal: 1/8” or 1/4” VCR, 3.2mm UPG

See Model Code

{C25D426D-36F2-421B-9D45-1B3453009219}
Electrical Connection

IP40: 15 pin D-Type connector

{B2061F4B-2841-4477-8384-B38AC14AC38A}
Electrical
{9B2C0B35-B3A4-4881-B327-251E06F8CCE6}
Analog Communication

0-5 Vdc, 0-10 Vdc, 0-20 mA, 4-20 mA

{4C0F2FAE-C938-4FB9-8A3B-EC1C630665D5}
Digital Communication

RS-485 (Smart Protocol)

{5EA92797-F284-4C28-A5BB-C1D75647400F}
Ratings
{CA2E0944-C762-450E-9161-FCBA17786D0B}
Temperature Range
0 - 60oC
{9E96C751-492D-4330-8D75-8634F1549D20}
Max Pressure

Standard Internal Valve: 500 psig

Optional Internal Valve: 1500 psig

{5DCD05AB-8D57-421B-9D31-8031C5BB3F48}
Power Supply / Consumption

Voltage: +14 - 27 Vdc

Nominal Current: 300 - 300 mA (controller); 100 - 150 mA (meter)

Maximum Current: 715 mA @ 14 Vdc (controller); 470 mA @ 14 Vdc (meter)

Maximum Power: 10.0 W (controller); 6.6 W (meter)

{BF7DCAAD-8DE1-475C-BCC3-9F80546DD2D7}
Diagnostics
{6D1F6413-977F-4939-BDE1-C6F5E3249A73}
Diagnostic Capability

LED Status light

Open Collector Alarm Output

Advanced diagnostics, alarms & warnings available via digital I/O

QMC QMB
Differentiator Third generation Coriolis sensor and electronics provides superior performance and reliability Large installed base with proven long-term performance
Response Time

SS Sensor Controller: <2 secs

SS Sensor Meter: <0.5 secs

SS Sensor Controller: <2 secs

SS Sensor Meter: <0.5 secs

Alloy C-22 Sensor Controller: <12 secs

Alloy C-22 Sensor Meter: <0.5 secs

Density Range 0 to 2.0 g/cc 0 to 0.3 and 0.5 to 2.0 g/cc
Wetted Materials

316L, 316L VAR, High Alloy Ferritic Stainless & 17-7PH

316L, 316L VAR, High Alloy Ferritic Stainless & 17-7PH

Optional: Alloy C-22 Sensor Tube

Housing

Standard: IP40

Standard: IP40

Optional: IP66

Optional: IP66XP

Fittings

Elastomer Seal: 1/16”, 1/8”, 1/4” or 6mm tube compression

1/8” or 1/4” NPT(F)

1/8” or 1/4” VCR, 1/4” VCO

Metal Seal: 1/8” or 1/4” VCR, 3.2mm UPG

See Model Code

1/16”, 1/8”, 1/4” or 6mm tube compression, VCR, VCO or NPT(F), 3.2mm UPG, Downport ANSI/ISA 76.00.02

See Model Code

Electrical Connection

IP40: 15 pin D-Type connector

IP40: 15 pin D-Type connector

IP66: Unpluggable Terminal Block 28-16 Awg

IP66XP: 3/4" NPT wiring access to IP40 device with 15 pin D-Type connector

Analog Communication

0-5 Vdc, 0-10 Vdc, 0-20 mA, 4-20 mA

0-5 Vdc, 4-20 mA

Digital Communication

RS-485 (Smart Protocol)

HART over 4-20 mA

Max Pressure

Standard Internal Valve: 500 psig

Optional Internal Valve: 1500 psig

Standard Internal Valve: 500 psig

Optional Internal Valve: 1500 psig

Optional Internal Valve: 4500 psig (Remote Valve for Controller)

Diagnostic Capability

LED Status light

Open Collector Alarm Output

Advanced diagnostics, alarms & warnings available via digital I/O

LED Status light

Open Collector Alarm Output

Filename File Type Documents By Language
Data Sheet - QMC pdf English
Quick Start Guide - QMC pdf English
Installation Manual - QMC pdf English
RS485 Supplemental Manual - QMC pdf English
Data Sheet - QMB pdf English
Quick Start Guide - QMB pdf English
Installation Manual - QMB pdf English
HART® Supplemental Manual pdf English

Use the following guide (model code information from data sheet) to translate the CAD configurations available below. If you need assistance or can’t find what you’re looking for, please contact Applications Engineering or your local Brooks Instrument representative.

Example: QMC - C - A - A - 1L - 1 - A

 Meter Series
 Type  Valve Type
 Enclosure
  Process Connections  Electrical Connection  Inline Filter
 QMC
 C - Controller  A - No Valve
 A - NEMA 1/ IP40
  1D - 1/8" Tube Compression Fittings  1 - 15 pin D-type   A - None
   M - Meter  B - Normally Closed
 
  1L - 1/8" VCR
  C - Inline Filter, Cartridge Filter, 20 Micron
 
 C - Remote Normally Closed
 



 

Access our offering of cables to connect our products to your process.
0250 Series
Controls and powers up to 4 devices
Brooks Expert Support Tool (BEST)
A Windows™ based application providing expanded control of Brooks Instrument MFCs and Pressure Controllers.